10
Operating Instructions Combics CAIXS2
Getting Started
Instead of an analog/digital converter (ADC), you could also install a data interface
to connect an intrinsically safe digital weighing platform or balance (e.g. an
IS......-X).
Connection using RS232 (Option A16)
1 CTS
green
2 DTR
brown
3 RxD
yellow
4 TxD
white
5 GND
gray
6 GND
Connection using RS485 (Option A19)
1 RxD-TxD-P white
2 RxD-TxD-N yellow
3 GND
gray
4 GND
3
Insulate unused cable ends
Connecting Intrinsically Safe Data Cables
Connect intrinsically safe data cables to COM 1 (RS232, RS485 or RS422 and
intrinsically safe control signals). For pin assignments on the data interface board,
see 66015-741-50.
External voltage supply for
AP1
At AP1 connected IS platform requires an external
Voltage supply, see document 66015-741-50 Page 2/9.
1
2
3
4
5 6
1
2
3
4
5 6
1
2
3
4
Summary of Contents for Combics 2 CAIXS2
Page 120: ...120 Operating Instructions Combics CAIXS2 Declarations of Conformity...
Page 121: ...Operating Instructions Combics CAIXS2 121 Declaration of Conformity...
Page 123: ...Operating Instructions Combics CAIXS2 123...
Page 149: ...Operating Instructions Combics CAIXS2 149...
Page 172: ...172 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 173: ...Operating Instructions Combics CAIXS2 173 IECEx Certificate...
Page 174: ...174 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 175: ...Operating Instructions Combics CAIXS2 175 IECEx Certificate...
Page 176: ...176 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 178: ...178 Operating Instructions Combics CAIXS2...