Specifications
Specifications
ADC Scale Interface
When used in standard applications (as opposed to legal metrology):
– Display resolution
≤
62500 d
– Using the equipment in legal metrology:
Load cell connection:
– Supply voltage
2,5 V
– Bridge impedance
83
Ω
to 2000
Ω
– Available sensor technology
4-conductor or 6-conductor technology
When used in legal metrology:
– Available sensor technology
6-conductor technology
– Max. cable length per gage
150 m/mm
2
– Lowest permissible input signal
for Pind = 0.5
0.2 µV/e
– Fraction of tolerance for this module:
for Delta U
min
0.2 µV/e
0.5
Measurement signal
0 mV to 7.5 mV
“Lowest permissible input signal
when used in non-legal metrology":
0.02 µV/d
Measurement signal for dead load
0 mV to 2.5 mV
Sensitivity
4 million digits max. (internal)
Digital protective interface
According to EN 45501
Data interface
Bidirectional intrinsically safe RS-232 interface “COM 1” with intrinsically
safe control outputs (digital I/Os) for connection to suitable intrinsically safe
equipment. Alternatives: Bidirectional intrinsically safe RS-422 interface
“COM 1” or bidirectional intrinsically safe RS-485 interface “COM 1” with
intrinsically safe control outputs (digital I/Os).
Additional data interface:
Optional
Display
20 mm LCD, 14-segment plus status symbols, backlit
Housing: – Material
Stainless steel 1.4301
–
Protection class according to EN 60529 IP69
Temperature range
Storage temperature -20°C to +60°C, operating temperature -10°C to +40°C
Power supply
Only via suitable and where applicable country-specific EX power supply
provided by Sartorius model YPS02-X.. / YPS02-Z.. / YPSC01-X / YPSC01-Z:
100-240Vac (± 10%), 50/60Hz; max. 25VA or 40-80VA with YPSC01-.
or via Ex battery pack YRB02-X
Explosion protection
See EC Type Examination Certificate in the Appendix
Emissions
In accordance with EN61326-1:2006 (IEC61326-1: Class A
Defined immunity to interference
In accordance with EN61326-1:2006 (IEC61326-1): Industrial areas
Electrical safety
In accordance with EN61010-1 (IEC61010-1)
Summary of Contents for Combics 2 CAIXS2
Page 120: ...120 Operating Instructions Combics CAIXS2 Declarations of Conformity...
Page 121: ...Operating Instructions Combics CAIXS2 121 Declaration of Conformity...
Page 123: ...Operating Instructions Combics CAIXS2 123...
Page 149: ...Operating Instructions Combics CAIXS2 149...
Page 172: ...172 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 173: ...Operating Instructions Combics CAIXS2 173 IECEx Certificate...
Page 174: ...174 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 175: ...Operating Instructions Combics CAIXS2 175 IECEx Certificate...
Page 176: ...176 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 178: ...178 Operating Instructions Combics CAIXS2...