Operating Instructions Combics CAIXS2
Configuring Weighing Platforms
Configuring Weighing Platforms
Service mode
Purpose
The Service mode enables access to additional menu items in the Setup menu
(
setup
) which are not displayed when the Service mode is not active. The most
important calibration and adjustment work for the indicator and for the connected
weighing platform can be carried out in the Service menu, e.g. ADC configuration.
When the Service mode is active, an “
S
" is shown in the top right-hand corner of
the display. To deactivate the Service mode, restart the indicator (turn the indicator
off and back on again).
In Service mode, the
Setup
menu is expanded with the following parameters after
entering the user password:
–
S-DATE
for entering the next service date
–
SER.No.
for entering the device serial number
–
MODEL
with the
model description
–
S-SQMIN
–
ALIB.MEM
for deleting the
alibi memory.
The Setup menu for
WP1
and
WP2
can be extended to include the following setting
options to confugure the weighing platforms:
Param1
cal.adj
calibration, adjustment
1.9
lIN.INT
internal linearization
1.9.5
lIN.ext
.
external linearization with default weights
1.9.6
lIN.e.usr
external linearization with user-defined weights (entered under 1.18)
1.9.7
SET.PREL.
set the preload (not for use in legal metrology)
1.9.8
DEL.PREL
delete the preload (not for use in legal metrology)
1.9.9
HND.XT.G/CAL.ADJ
enter the adjustment and linearization weights
1.18
CAL.ADJ.1
enter adjustment weight
1.18.1
LIN. WT.1
enter lin. weight 1
1.18.2
LIN. WT.2
enter lin. weight 2
1.18.3
LIN. WT.3
enter lin. weight 3
1.18.4
LIN. WT.4
enter lin. weight 4
1.18.5
ADJ.W/O.W
adjust without weights (enter the characteristic data of the load cells)
1.19
NOM.LOAD
. nominal load
1.19.1
RESOLUT
resolution (only visible for older ADCs)*
1.19.2
SENSIT.1
sensitivity in mV/V for cell 1 (or average value for all cells)
1.19.3
SENSIT.2
sensitivity in mV/V for cell 2
1.19.4
SENSIT.3
sensitivity in mV/V for cell 3
1.19.5
SENSIT.4
sensitivity in mV/V for cell 4
1.19.6
ZER.POIN
zero point or offset of system in mV/V
1.19.7
SAVE.
save values for 1.19
1.19.8
GEOG.DAT
adjustment location (geograph. data; or alternatively the gravitational acceleration
at the place of installation)
1.20
LATITUD
latitude in degrees
1 20.1
ALTITUD
elevation in meters above sea level
1 20.2
GRAVITY
. gravitational acceleration
1 20.3
SAVE
. save values for 1.20
1 20.4
ADC settings (see Analog/Digital Converter)
11
Apply the serial number of the IS weighing platform
12.1
Apply the serial number
12.1.1
Inactive (standard WP)
12.1.2
Summary of Contents for Combics 2 CAIXS2
Page 120: ...120 Operating Instructions Combics CAIXS2 Declarations of Conformity...
Page 121: ...Operating Instructions Combics CAIXS2 121 Declaration of Conformity...
Page 123: ...Operating Instructions Combics CAIXS2 123...
Page 149: ...Operating Instructions Combics CAIXS2 149...
Page 172: ...172 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 173: ...Operating Instructions Combics CAIXS2 173 IECEx Certificate...
Page 174: ...174 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 175: ...Operating Instructions Combics CAIXS2 175 IECEx Certificate...
Page 176: ...176 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 178: ...178 Operating Instructions Combics CAIXS2...