Operating Instructions Combics CAIXS2
69
Operation
Checkweighing
H
With this application, you can check whether the sample on the weighing platform
matches a target value or lies within a given tolerance range. Checkweighing also
makes it easy to fill sample materials to a specified target weight.
Selection and settings in the
Applic.
/ Applic.2/
CHECK.WG
menu.
Features
– Enter the nominal or target weight (set point) and the tolerance range
delimiters either using the keypad or by saving the weight value of a load on
the platform.
– Enter the tolerance limits as absolute values (Min and Max), as a percentage
deviation from the target or as a relative deviation from the target.
Setting:
Applic./Applic.2/CHECK.WG/CHECK.RG
menu item 4.5.
– The target value can be taken over as a weighed value from a weighing
platform, and the upper and lower tolerance limits are defined as a percentage
deviation from the target value (setting 4.5.2). The following percentages can
be selected as the deviation: 0.1%, 0.2%, 0.5%, 1%, 1.5%, 2%, 3%, 5% or 10%,
selection using the
r
key.
– The target value, lower tolerance limit (minimum) and upper tolerance limit
(maximum) can applied as weighed values from the weighing platform
(menu item 4.5.1).
– The target value can be applied as weighed values and via asymmetrical percent
limits (menu item 4.5.3).
– The target value can be applied as weighed values and via relative weight limits
(menu item 4.5.4).
– Target value and tolerance limits checked during input; values must conform
to: upper limit > target > lower limit > 1 digit.
– Checkweighing range: either 30% to 170% of the target, or from 10% to
infinity.
– Results are shown on the main display, as a bar graph and LED as well as sent
to control output ports for further processing.
– Toggle the main display between weight and tolerances limits by pressing the
w
key. If the weight in the readout is outside the tolerance range, “LL" (too
low) or “HH" (too high) is displayed.
– Activate Info mode with the
I
key.
– Automatic results printout (
Applic./Applic.2/CHECK.WG/CHECK.RG
menu
item 4.6).
– Automatic taring of container weight (
APPLIC./ aut.tare
menu item 3.7).
– Automatic initialization when you switch on the scale with most recently saved
application data (
APPLIC./aut.Strt
menu item 3.8).
You can assign different functions to the
c
key to delete applications. When you
clear applications, you can delete either the data stored for all applications or just
selected data stored for the active application.
Setting:
APPLIC/CLER.CF
menu item 3.24.
Tare Function:
If you store a tare (weight value) by pressing the
)
key, you can later enter a tare
value manually. The tare value you enter is added to the stored tare value.
Setting:
APPLIC./TARE.FNC
menu item 3.25.1 (factory setting).
Summary of Contents for Combics 2 CAIXS2
Page 120: ...120 Operating Instructions Combics CAIXS2 Declarations of Conformity...
Page 121: ...Operating Instructions Combics CAIXS2 121 Declaration of Conformity...
Page 123: ...Operating Instructions Combics CAIXS2 123...
Page 149: ...Operating Instructions Combics CAIXS2 149...
Page 172: ...172 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 173: ...Operating Instructions Combics CAIXS2 173 IECEx Certificate...
Page 174: ...174 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 175: ...Operating Instructions Combics CAIXS2 175 IECEx Certificate...
Page 176: ...176 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 178: ...178 Operating Instructions Combics CAIXS2...