78
Operating Instructions Combics CAIXS2
Operation
By
entering a percentage
: The upper value of Class 1 is entered using the keypad
or by saving the value indicated. Upper limits for the other classes are defined by
entering a percentage of deviation from the upper limit of Class 1, using the
keypad.
Example
: Enter 100 g as the upper limit of Class 1. Then enter 15%. When working
with 3 classes, this yields the following weight classes:
Class 0: up to the minimum load
Class 1: >minimum load - 100 g
Class 2: >100 g – 115 g
Class 3: >115 g - maximum load
When working with 5 classes, this yields the following weight classes:
Class 0: up to the minimum load
Class 1: >minimum load - 100 g
Class 2: >100 g – 115 g
Class 3: >115 g – 130 g
Class 4: >130 g – 145 g
Class 5: >145 g - maximum load
The values entered remain valid until deleted by pressing the
c
key or until
overwritten by a new value. They remain saved after the scale is switched off.
Preparation
t
Open the
APPLIC./APPLIC.2/CLASS
menu.
Available parameter settings
* = Factory setting
MIN.INIT
Minimum load for initialization
3.6
1 digit
1 scale interval
3.6.1*
2 digit
2 scale intervals
3.6.2
5 digit
5 scale intervals
3.6.3
10 dig.
10 scale intervals
3.6.4
20 dig.
20 scale intervals
3.6.5
50 dig.
50 scale intervals
3.6.6
100 dig.
100 scale intervals
3.6.7
200 dig.
200 scale intervals
3.6.8
500 dig.
500 scale intervals
3.6.9
1000 d
1000 scale intervals
3.6.10
CTRL.SET
Activate SET control output
4.3.
OUTPUT
“SET" output
4.3.1*
OP.READY
Ready to operate
4.3.2
OUTP.ACT
port lines
4.7
OFF
Off
4.7.1*
ALWAYS
Always on
4.7.2
STABL.
On at stability
4.7.3
QTY.
Number of classes
4.8.
3 CLASS
3 classes
4.8.1*
5 CLASS
5 classes
4.8.2
INPUT
Parameter input
4.9.
WEIGHTS
Weight values
4.9.1*
PERC.TAG
Percentage values
4.9.2
Automatic printing
4.10.
MANUAL
Off
4.10.1*
automat
On
4.10.2
)
t
To save the setting, press the
)
key.
( (
...
t
To exit setup: Press the
(
key several times.
Summary of Contents for Combics 2 CAIXS2
Page 120: ...120 Operating Instructions Combics CAIXS2 Declarations of Conformity...
Page 121: ...Operating Instructions Combics CAIXS2 121 Declaration of Conformity...
Page 123: ...Operating Instructions Combics CAIXS2 123...
Page 149: ...Operating Instructions Combics CAIXS2 149...
Page 172: ...172 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 173: ...Operating Instructions Combics CAIXS2 173 IECEx Certificate...
Page 174: ...174 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 175: ...Operating Instructions Combics CAIXS2 175 IECEx Certificate...
Page 176: ...176 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 178: ...178 Operating Instructions Combics CAIXS2...