148
Operating Instructions Combics CAIXS2
Verification of Intrinsic Safety
Verification of
Intrinsic Safety
CAIXS2 /
Rev. 00
Klausgrete
2013-08-08
Sheet
of
8
8
66015-751-60
RS422 Data Interface
Non-hazardous
area
Hazardous area
YCO01-Y
Interface converter
Note:
1: Per circuit
2: Combined circuits
*:
Versus GND; **: Between the lines; ***: Resistively limited
Cable:
For a standard cable (e.g., type LiYCY with max. 250nF/km
(wire versus wire or wire versus shield), max. 750µH/km and
at least 34 ohms/km (e.g. 0.5mm² stranded wire), this yields
Lo/Ro = 22 µH/ohm Permitted length of the flexibly installed
cable:
330m
YDI05-Z
(Option A25)
Interface converter
II (2) GD [EEx ib] IIC
Alternative
connection
RS422 data interface
2)
(
COM1
)
Uo
5,2 V
Ui
8.6 V
Io
290 mA
Ii
210 mA
Po
496 mW
Pi
0.5 W
Co
60 µF
Ci
0.5 µF
Lo
300 µH
Li
0 mH
Lo/Ro 50 µH/
RS422 data interface of YCO01-Y
2)
Uo
7.2 V*
Ui
12.6 V*
8.0 V**
25.2 V**
Io
207 mA***
Ii
1.5 A
Po
330 mW
Pi
2.5 W
Co
13.5 µF*
Ci
1 nF
8.4 µF**
Lo
0.7 mH*
Li
1.6 µH
0.7 mH**
Lo/Ro 36 µH/ */ 36µH/ *
RS422 data interface of YDI05-Z
2)
Uo
6.0 V *
)
Ui
12.6 V *
)
6.8 V **
)
25.2 V **
)
Io
172 mA (linear) Ii
0.2 A
Po
0.5 W
Pi
2.3 W
Co
12 µF *
)
Ci
28 µF *
)
Co
17.7 µF
**)
Ci
200 nF **
)
Lo
0.7 mH
Li
2 µH
CAIXS2
II 2 G Ex ib IIC T4 Gb
II 2 D Ex ib IIIC T80°C Db
Zone 1, 2, 21, 22
Gas: IIC T4
Dust: IIIC
Verification of
Intrinsic Safety
CAIXS2 /
Rev. 00
Klausgrete
2013-08-08
Sheet
of
7
8
66015-751-60
RS485 Data Interface
Non-hazardous area
Hazardous area
Note:
1: Per circuit
2: Combined circuits
4: BAS01ATEX7005; II (1) GD [EEx ia] IIC
5: Data for CIXS3
6:
Data for
CAIXS2, SIWXS…-3-..-... and ISX...-..-...
7: Both channels connected on Z966
8: Only 2 RS232 connections are used on the YDI05-Z
*: Versus GND; **: Between the lines; ***: Resistively limited
YDI05-Z.. interface converter
II (2) GD [EEx ib] IIC
or
Z966 Zener barrier
4)
(in YDI01-Z) II (2) G [EEx ib] IIC
or
YCO01-Y interface converter
II (2) GD [EEx ib] IIC
or
II 3 (2)GD EEx nR[ib]IIC T4
Up to 7 additional
CIXS3
or
CAIXS2 /
SIWXS…-3-..-... /
ISX...-..-...
3)
Data cable:
Recommended: Sartorius cable
YCC485-X
with approx.
10µH/ohm and 120pF/m (wire/wire)
up to 1000m flexibly installed
.
Z966
1)
pins A/J/K/N and M
Uo
12 V*
24V**
Io
82 mA /164mA
7)
Po
0.24 W / 0.48W
7)
Co
1.41 µF*
125nF**
Lo
5.52 mH
1.38mH
6)
Lo/Ro 147 µH/
57 µH/
6)
YDI05-Z..
2) 8)
Uo
12.4 V*
Ui
12.6 V*
24.8V**
25.2 V**
Io
130 mA***
Ii
any
Po
400 mW
Pi
any
Co
1.24 µF*
Ci
0
112nF**
Lo
0.4 mH*
Li
0
0.4 mH**
Lo/Ro 44 µH/ * / 22 µH/ **
YCO01-Y
2)
Uo
7.2 V*
Ui
12.6 V*
8.0 V**
25.2 V**
Io
207 mA***
Ii
1.5 A
Po
330 mW
Pi
2.5 W
Co
13.5 µF*
Ci
1 nF
8.4 µF**
Lo
0.7 mH*
Li
1.6 µH
0.7 mH**
Lo/Ro 36 µH/ * / 36µH/ **
RS485 data interface
2) 6)
Ui see below
Uo
5.2 V
Ii see bwlow
Io
210 mA***
Pi any
Po
263 mW
Ci 260 nF
Co
60 µF
Li 0 mH
Lo
600 µH
Lo/Ro 125 µH/
Ui
±
12.4V
12.0V
7.2V
Ii 130 mA***
164mA*** any
RS485 data interface
5) 2)
(
LV4
)
Uo
7.2 V
3)
Ui 12.6 V
8.2 V
4)
Io
168 mA*
Ii
1.5 A*
Po
0.25 W
Pi
2.5 W
Co
13 µF
3)
Ci 300 nF
3)
Co
7.6 µF
4)
Ci 100 nF
4)
Lo
0.8 mH
Li
0 mH
Lo/Ro 118 µH/ohm
RS485 data interface
5) 2)
(
UNICOM
.
LV2
)
Uo
7.2 V
Ui 12.6 V
Io
127 mA*
Ii
1.5 A*
Po
0.273 W
Pi
2.5 W
Co
11.3 µF
Ci 0
Lo
2 mH
Li
2 µH
Lo/Ro 118 µH/ohm
CAIXS2
II 2 G Ex ib IIC T4 Gb
II 2 D Ex ib IIIC T80°C Db
Zone 1, 2, 21, 22
Gas: IIC T4
Dust: IIIC
Summary of Contents for Combics 2 CAIXS2
Page 120: ...120 Operating Instructions Combics CAIXS2 Declarations of Conformity...
Page 121: ...Operating Instructions Combics CAIXS2 121 Declaration of Conformity...
Page 123: ...Operating Instructions Combics CAIXS2 123...
Page 149: ...Operating Instructions Combics CAIXS2 149...
Page 172: ...172 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 173: ...Operating Instructions Combics CAIXS2 173 IECEx Certificate...
Page 174: ...174 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 175: ...Operating Instructions Combics CAIXS2 175 IECEx Certificate...
Page 176: ...176 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 178: ...178 Operating Instructions Combics CAIXS2...