Operating Instructions Combics CAIXS2
147
Verification of Intrinsic Safety
Verification of
Intrinsic Safety
CAIXS2 /
Rev. 00
Klausgrete
2013-08-08
Sheet
of
6
8
66015-751-60
RS232 Data Interface
Non-hazardous area
Hazardous area
RS232 data interface
(pins A/J/K/N and M)
1)
Ui
12.6 V
Uo
12.6 V
Ii
85 mA
Io
28 mA
Pi
270 mW
Po
88 mW
Ci
3 nF
Co
1.15 µF
Li
0
Lo
50 mH
Note:
1: Per circuit
2: Combined circuits
3: Not all models can be used in zone 21,22
4: BAS01ATEX7005; II (1) GD [EEx ia] IIC
5: Both channels connected on Z966
*: Versus GND; **: Between the lines; ***: Resistively limited
YDI05-Z.. interface converter
II (2) GD [EEx ib] IIC or
Z966 Zener barrier
4)
in YDI02-Z..: II (2) G [EEx ib] IIC or
YCO01-Y interface converter
II (2) GD [EEx ib] IIC
or
II 3 (2)GD EEx nR[ib]IIC T4
6-wire standard cable, up to 0.5mm² stranded wire, with
max. 250nF/km, 750µH/km and minimum 34ohm/km has
22µH/ohm. However, the length of the cable (
flexibly
installed
) is limited to
under 25m
by the RS-232
specifications. 6-wire cable, type PR6136 (Lmax =
1.1mH/km; Cmax = 220nF/km; Rmin = 26 ohms/km:
43µH/ohm): under 25m (see above).
Alternative
connections
YDI05-Z..
2)
Uo
12.4 V*
Ui
12.6 V*
24.8 V**
25.2 V**
Io
260 mA*** Ii
any
Po
800 mW
Pi
any
Co
1.24 µF*
Ci
0
112 nF**
Lo
400 µH*
Li
0
400 µH**
Lo/Ro 44 µH/ * / 22 µH/ **
Z966
1)
pins A/J/K/N and M
Uo
12 V*
24 V**
Io
82 mA /164 mA
8)
Po
0.24 W / 0.48 W
8)
Co
1.41 µF*
125 nF**
Lo
5.52 mH
1.38 mH
8)
Lo/Ro 147 µH/
57 µH/
8)
YCO01-Y
2)
Uo
11.8 V*
Ui 12.6 V*
23.6V**
25.2 V**
Io
123 mA***
Ii
131 mA
Po
361 mW
Pi
any
Co
1.5 µF*
Ci 0.5 nF
129 nF**
Lo
2 mH*
Li
0.8 µH
2 mH**
Lo/Ro 98 µH/ * / 98µH/ **
SIWS… / ISXS.... / CAIXS2 RS232 data interface
2)
COM1
Ui
12.6 V* / 25,2V**
Uo
10 V* / 20V**
Ii
328 mA***
Io
101 mA***
Pi
any
Po
253 mW
Ci
2.2 nF* / 0.5nF**
Co
3 µF* / 217nF**
Li
0 mH
Lo
3 mH
Lo/Ro
140 µH/
CAIXS2
II 2 G Ex ib IIC T4 Gb
II 2 D Ex ib IIIC T80°C Db
Zone 1, 2, 21, 22
Gas: IIC T4
Dust: IIIC
FC/FCA/IS.......-.X..
3)
scale / weighing platform
SIWXS…-3-..-... /
ISX...-..-... / CAIXS2
3)
scale / weighing platform
Verification of
Intrinsic Safety
CAIXS2 /
Rev. 00
Klausgrete
2013-08-08
Sheet
of
5
8
66015-751-60
Digital I/Os
Non-hazardous area
Hazardous area
Safety barrier
1)
Uo
8.6 V
Io
50 mA
Po
107.5 mW
Co
6.2 µF
Lo
15 mH
Foot switch
3)
9001/01-086-050-101
Safety barrier
(supplied by Stahl)
Zone 1, 2, 21, 22
Gas: IIC T4
Dust: IIIC
Junction box
3)
Digital I/Os (COM1)
2)
Ui
8.6 V
Uo
6,0 V
Ii
--
Io
45 mA
Pi
--
Po
67 mW
Ci
0 nF
Co
40 µF
Li
0 mH
Lo
20 mH
Lo/Ro 530 µH/
Switch
3)
Alternative
connection
Cable with up to 10 wires and with max.
250nF/km and 1.1mH/km: maximum cable
length 1000m (flexibly installed).
Note:
1: Per circuit
2: Combined circuits
3: Suitable equipment may be optionally used; only passive wiring/switch
CAIXS2
II 2 G Ex ib IIC T4 Gb
II 2 D Ex ib IIIC T80°C Db
Summary of Contents for Combics 2 CAIXS2
Page 120: ...120 Operating Instructions Combics CAIXS2 Declarations of Conformity...
Page 121: ...Operating Instructions Combics CAIXS2 121 Declaration of Conformity...
Page 123: ...Operating Instructions Combics CAIXS2 123...
Page 149: ...Operating Instructions Combics CAIXS2 149...
Page 172: ...172 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 173: ...Operating Instructions Combics CAIXS2 173 IECEx Certificate...
Page 174: ...174 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 175: ...Operating Instructions Combics CAIXS2 175 IECEx Certificate...
Page 176: ...176 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 178: ...178 Operating Instructions Combics CAIXS2...