106
Operating Instructions Combics CAIXS2
Data interfaces
Configuring a Printout
Printouts are configured in the
Setup
menu under “Printouts" (
setup / print /
protoc.
). This should be carried out
after
configuring the application since some
data in the printout is application-dependent.
You can configure a separate printout for each interface. Each printout is comprised
of different information blocks that can be activated or deactivated via multiple
selection in the menu.
For the “Totalizing" and “Net-total Formulation" applications, the totalizing/results
printout can be configured independent of the individual/component printout.
Headers
2 headers each with a max. of 20 characters are available (e.g., for printing the
company name).
Input: menu items 7.4.1 and 7.4.2. Empty headers are not printed.
Example
Print image:
EISENSCHMIDT
GOETTINGEN
In this example, the company name is printed centered because there are 4 and 5
spaces before the text.
GMP-Compliant Printouts
When this function, the printout is supplemented with a GMP header and a GMP
footer (GMP: “Good Manufacturing Practice”).
Setting: menu item 7.13.
The GMP header precedes the first measured result. The GMP footer is printed
either after each individual measurement result ("GMP-compliant printout always
for 1 result", 7.13.2) or after the last result in a series of measurements ("GMP-
compliant printout always for several application results,” menu item 7.13.3).
To end a series of measured results, press and hold the
p
key. In this case,
the
p
symbol is displayed after the GMP header is printed and remains in the
display until the GMP footer is printed.
If you toggle to a different platform while a GMP printout of several measured
results is being generated (7.13.3), the GMP footer for the platform used up to
that point is generated when you press the
n
key. The GMP header for the other
platform is included on the next printout generated.
A GMP-compliant printout is generated automatically at the conclusion of
calibration/adjustment, linearization routines, as well as when you set or clear a
preload.
When printing GMP-compliant printouts on label printers under menu setting
7.13.3, the relationship between the GMP header and footer is lost (printed on
different labels). GMP-compliant printouts on label printers, therefore, should only
take place using menu setting 7.13.2. Three examples of GMP headers and one
example of a footer are shown in the following.
Summary of Contents for Combics 2 CAIXS2
Page 120: ...120 Operating Instructions Combics CAIXS2 Declarations of Conformity...
Page 121: ...Operating Instructions Combics CAIXS2 121 Declaration of Conformity...
Page 123: ...Operating Instructions Combics CAIXS2 123...
Page 149: ...Operating Instructions Combics CAIXS2 149...
Page 172: ...172 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 173: ...Operating Instructions Combics CAIXS2 173 IECEx Certificate...
Page 174: ...174 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 175: ...Operating Instructions Combics CAIXS2 175 IECEx Certificate...
Page 176: ...176 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 178: ...178 Operating Instructions Combics CAIXS2...