Configuring Weighing Platforms
Minimum load (min. load)
When “Standard configuration” is used, this menu item is not displayed.
The minimum load of the connected weighing platform is entered under this menu
item. The minimum load for scales of class
l
is 20 e and 10 e for class
m
.
Attention:
The function of the minimum load setting is to warn operators that
below this limit, the summation of tolerances might lead to significant measure-
ment errors. In Germany, for example, initial weights below the minimum load are
not allowed.
Available weight units
WT.UNIT
Menu item 11.7
This menu item is used to select the weighing units that have been cleared for use
in weighing. All units marked with a circle (o) have been cleared for use, multiple
selection is possible.
Calibration/Adjustment unit
Cal.Unit
Menu item 11.8
This menu item is used to select the weighing unit that must be used for a
calibration/adjustment. The selected unit is then valid as a calibration/adjustment
unit even when a different unit is used during normal weighing operation.
Save parameters
SAVE
Menu item 11.10
The ADC configuration data is saved by selecting Menu item 11.10.1.
Summary of Contents for Combics 2 CAIXS2
Page 120: ...120 Operating Instructions Combics CAIXS2 Declarations of Conformity...
Page 121: ...Operating Instructions Combics CAIXS2 121 Declaration of Conformity...
Page 123: ...Operating Instructions Combics CAIXS2 123...
Page 149: ...Operating Instructions Combics CAIXS2 149...
Page 172: ...172 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 173: ...Operating Instructions Combics CAIXS2 173 IECEx Certificate...
Page 174: ...174 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 175: ...Operating Instructions Combics CAIXS2 175 IECEx Certificate...
Page 176: ...176 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 178: ...178 Operating Instructions Combics CAIXS2...