E 7.2 Display exchange
In case of a faulty display the complete mounting unit with display, pcb, keys
and inverter has to be exchanged. That is why only a precast mounting unit
is available as a spare part (no. 129691). Additionally a new scale is required
(spare part no. 167217). Please send back the faulty unit to RTW.
Proceed as follows to exchange the display:
Note:
When servicing it is very important to observe all standard ESD (electro-
static discharge) protection procedures.
1. For opening the unit please follow steps 1. to 3. described in Chapter E 7.1
2. Carefully remove the flat display connecting cable from the connector socket
right above on the main pcb (A). Avoid damage of the flat cable and the
connector. (Fig. E 7-3)
Fig. E 7-3: Loosen the connecting cable
E 7 Service E 7.2 Display exchange
Summary of Contents for 10689
Page 7: ...DigitalMonitor 8ch 10689 Be isp iel 10 68 9 P LU S Bedienungsanleitung deutsch...
Page 8: ...RTW DigitalMonitor 8ch 10689 8...
Page 14: ...RTW DigitalMonitor 8ch 10689 14...
Page 22: ...RTW DigitalMonitor 8ch 10689 22...
Page 50: ...RTW DigitalMonitor 8ch 10689 50...
Page 68: ...RTW DigitalMonitor 8ch 10689 68...
Page 74: ...RTW DigitalMonitor 8ch 10689 74...
Page 90: ...RTW DigitalMonitor 8ch 10689 90...
Page 94: ...RTW DigitalMonitor 8ch 10689 94 Anhang A Technische Daten...
Page 96: ...RTW DigitalMonitor 8ch 10689 96...
Page 99: ...DigitalMonitor 8ch 10689 Ex am ple 10 68 9 P LU S Operating Manual english...
Page 100: ...RTW DigitalMonitor 8ch 10689 100...
Page 106: ...RTW DigitalMonitor 8ch 10689 106...
Page 114: ...RTW DigitalMonitor 8ch 10689 114...
Page 138: ...RTW DigitalMonitor 8ch 10689 138...
Page 156: ...RTW DigitalMonitor 8ch 10689 156...
Page 162: ...RTW DigitalMonitor 8ch 10689 162...
Page 178: ...RTW DigitalMonitor 8ch 10689 178...
Page 182: ...RTW DigitalMonitor 8ch 10689 182...
Page 184: ...RTW DigitalMonitor 8ch 10689 184...