FAQ
49
21
.
There is foreign matter in the Air Compressor hoses. How could this happen?
Foreign matter can accumulate due to:
1. Reflux of materials inside the Air Compressor. This can be further prevented by making sure the pressure is properly emptied
using the pressure-vacuum valve after every use.
2. Condensation or moisture inside hoses: This can be resolved by removing moisture in the air filter.
To prevent foreign matter build-up, regularly check the drain container at the back of the Air Compressor and replace the air dry
filter once every year.
23.
What is the maintenance guide for the HEPA filter?
1. The HEPA filter should be replaced once every 6 months to 1 year. Consult the customer service / after-sales technicians.
2. If the end-user were to directly replace the filter, open the INVIVO upper lid and remove the fan and brackets first before
removing the filter. Assemble in the reverse order.
24.
What is the maintenance guide for the extruder nozzles?
1. If daily use, the extruder nozzles should be replaced in 3 months. Under normal circumstances, the period is 6 months.
2. Extruder heads should be replaced entirely, not just the nozzles.
22.
How do I use the regulators in the Air Compressor?
1. Regulators are used to control the amount of pressure delivery from the Air Compressor to the Air Dispenser Controller.
2. The Air Dispenser Controller set pressure will not exceed the maximum amount set with the regulators on the Air Compressor.
22
.
What is the maximum temperature and pressure that can be set on the Air Dispenser Controller?
1. The maximum temperature is 350℃.
2. The maximum pressure is 10 bar.