Dr. INVIVO 4D User Guide
Use of the Temperature Controller (Upgrade & Premium)
If there is no coolant in the equipment
1. Open the lid of the case and add coolant.
2. Make sure the valves of the Temperature Controller and Dr. INVIVO are open.
3. After plugging in the power, wait 10 seconds after the coolant starts flowing. Closer the coolant inlet with the provided silicone cover.
4. Refer to “Installation of the Temperature Controller” for more information.
24
1. The coolant will evaporate over time, so check regularly and refill.
2. The coolant supplied with Dr. INVIVO is a chemical. Be sure to read the precautions on the container surface and use them correctly.
3. The replacement term of Dr. INVIVO@COOLANT is 1 year. Using genuine coolant is recommended.
If there is coolant in the equipment
1. Make sure it has enough coolant. If not, turn off the power and then add coolant.
2. After setting the temperature of [BED, SYRINGE] of the parts to be used, press the [SET] button.
3. Do not turn off the power immediately after using the equipment. Set the temperature to 25 ℃ ~ 30 ℃, and turn off the power when
the temperature has been reached.
If the power is turned on when the coolant is not sufficiently circulated to the equipment, the temperature of the printing bed
and dispenser will continue to rise, which may affect internal element and cause a malfunction. Please be careful.
If the power is turned off without the above 3 steps, this may affect the internal device and cause malfunction.
Temperature
Set Temperature
Set Button
Control Arrows