Confidential & Proprietary|© ROKIT Healthcare Inc. 2019. All Rights Reserved
Dr. INVIVO 4D Installation
1. Unbox and place the Temperature Controller on a level and unclustered surface.
2. After placing each provided metal clamp on either end of the rubber hoses, connect the hoses between the Temperature Controller and the
Dr. INVIVO main body.
3. Clamp the rubber hoses and valves securely (to prevent leakage of the coolant).
4. Open the valves at the Temperature Controller and the Dr. INVIVO body.
5. Lift the silicone cover of the coolant inlet of the tank and pour 5 bottles of the coolant supplied with Dr. INVIVO (6 bottles are supplied in
total)
6. Close the coolant inlet with the silicone cover.
7. Insert the power plug into a grounded power outlet. Turn on the power switch of the Temperature Controller. Make sure that the cable is
not yet connected between the Temperature Controller and the Dr. INVIVO body.
8. Adjust the set temperature of the Temperature Controller to room temperature (20 ~ 25 ℃).
9. Circulate the coolant for 3 to 4 minutes until there are no bubbles in the hose.
10. After the coolant circulates, check the inside of the tank and add coolant if the level has dropped.
11. Turn off the power of the Temperature Controller.
12. Align the groove of temperature control cable with the one in the socket and connect it gently. Do not use excessive force.
13. Turn on the body and the Temperature Controller together. Operate the Temperature Controller to verify that the bed and syringe
temperature are adjustable. First test at -1 ℃, then test at 25 ℃.
14. Turn off the power and end the installation.
1. Make sure to follow the above instructions step by step. Only after the installation of the Temperature Controller is complete and the coolant has
been circulated sufficiently in the Dr. INVIVO body, turn on the Dr. INVIVO body.
2.
When connecting the electric system cable, connect only when the power is off.
3.
The coolant supplied with Dr. INVIVO is a chemical. Be sure to read the precautions on the container surface and use them correctly.
Installation of Temperature Controller (PREMIUM & UPGRADE)
[Packaging specification 520x480x365, 16.5 kg ]
1.
Use the original coolant supplied with Dr. INVIVO.
2.
The coolant will evaporate over time, so check regularly and refill.
CAUTION
WARNING
15
INV-CHS-20191210