Dr. INVIVO 4D User Guide
Bed Leveling – Right, Dispenser 1
1. Adjust the left nozzle so that its z-position is higher than the right nozzle. If the left nozzle is lower than the right nozzle, the left nozzle
may hit the bed and be damaged.
2. Select [New Creator K APP] > [Bed Leveling] > [RIGHT POSITION] on Dr. INVIVO display.
3. Select [HOME] button to move the Dr. INVIVO bed to the origin of the right nozzle set in 'Set X-Y Position'.
4. Select [HOME] button allowing an unlock of the z-axis limit sensor so that the nozzle can move closer to the bed.
This does not need to
be done from Firmware v1.70 and on.
5. Select [SET HOME POSITION], and select the desired travel distance among [0.1MM] [1MM] [10MM].
6. Select the arrows on the Z-axis up and down to adjust the gap between the nozzle tip and the bed.
7. Use the thickness gauge to adjust distance between the nozzle tip and the bed. Even if using the same nozzle diameter, the ideal gap
will vary across materials.
8. Select [FINISH] to finish the bed leveling process.
9. If you are leveling on the right side only, use the micrometer to lift the left nozzle away from the bed for safety.
1. Please click on the LCD display slowly. Clicking on LCD display repeated may lead to malfunction on the display.
2. Place a culture dish (or any other consumable to print on) on the bed before starting the bed leveling process.
3. Be careful not to hit the right / left nozzles (or needles) against the print bed.
4. Start the Bed Leveling only after confirming that right and left printheads and nozzles have been installed properly to the system,
5. Avoid touching the print bed with the thickness gauge. Please use the thickness gauge after sterilization.
33
INV-CHS-20191210
CAUTION