Dr. INVIVO 4D User Guide
Bed Leveling – Left, Dispenser 2
34
When Dual Printing with Left Dispenser
1. Place the leveling jig on the print bed.
2. Lower the right nozzle tip until it is just about to touch the leveling jig. Use the thickness gauge to adjust this distance to be about 0.1
mm.
3. Use the micrometer at the top left corner to lower the left nozzle tip until it is about to touch the leveling jig. Use the thickness gauge.
4. Once the right and the left nozzle tips have been aligned in the Z-axis with respect to the leveling jig, bed leveling is complete.
1. Please click on LCD button slowly. Clicking on LCD button continuously may lead to malfunction on the display.
2. Place printing bed (e.g, Peri Dish) before starting the bed leveling process.
3. Be careful not to hit the right / left nozzles (or needles) against the printing bed (e.g. Petri Dish).
4. Start the bed leveling after confirming that the syringe or barrel containing the printing materials on the right and left sides are correctly
installed.
5. Do not press the printing bed with the thickness gauge. Please use the thickness gauge after sterilization.
INV-CHS-20191210
CAUTION