Bed Leveling – Left, Dispenser 2
When Single-Printing with Left Dispenser
1. Adjust the left nozzle so that its z-position is higher than the right nozzle. If the left nozzle is lower than the right nozzle, the left nozzle
may hit the bed and be damaged.
2. Select [New Creator K APP] > [Bed Leveling] > [LEFT POSITION] on INIVIVO display.
3. Select [HOME] button to move the printing bed to the origin of the right nozzle set in 'Set X-Y Position'.
4. Select [HOME] button of z-axis. After selecting this button, nozzle can move below the z-axis limit sensor.
5. Select [SET HOME POSITION], and choose the travel distance among [0.1MM] [1MM] [10MM].
6. Select the arrows on the Z-axis up and down to adjust the gap between the nozzle tip and the bed.
7. Use the thickness gauge to adjust distance depending on the characteristic of materials.
8. Select [FINISH] to finish the bed leveling process.
9. Raise the nozzle up for safety.
35
INV-CHS-20191210
CAUTION
1. Beginning with Firmware v.1.70, there is no “Left Position” button. Use the same process used in dual printing to level the left dispenser.