Dr. INVIVO 4D User Guide
Preparation Before Use
1. Make sure that the Dr. INVIVO Body and accessories are installed and working properly.
2. Make sure that the Dr. INVIVO Body and accessories are set to the desired setting. It is recommended to warm up the system
30 minutes before printing.
3. Use the proper laboratory standards (i.e. wearing globes) to operate the system in an optimal condition.
4. Use the proper consumables provided by or confirmed with the manufacturer (i.e. barrels, syringes, nozzles, needles).
5. Check that each barrel, syringe, nozzle and needle are not deformed or contaminated. Replace or clean them if they are
deformed or contaminated.
6. Fill each barrel or syringe with the material to be printed.
7. Connect the appropriate nozzle or needle to the barrel or syringe for output.
8. Correctly install a genuine barrel or genuine syringe containing the material to be printed in the Dr. INVIVO Body.
9. If you want to run the output from a PC, install NewCreatorK on your PC and connect it with the Dr. INVIVO. You can
connect with a USB cable or Wi-Fi.
1. Do not overtighten the fixing screw when securing the syringe. The syringe may be broken or the cylinder may be deformed
and the X-Y-Z position may not be set correctly.
2. Do not exceed the capacity when filling the barrel or syringe with material.
3. When filling the barrel or syringe, be careful not to create air bubbles. Bubbles mixed with the material will degrade the
output quality.
4. Press the LCD button slowly. Pressing the LCD button repeatedly may cause a malfunction.
23
INV-CHS-20191210
CAUTION