System Overview
R&S
®
ZVA
87
Quick Start Guide 1145.1090.62 ─ 10
●
Imported and user-defined kits can be changed in the "Calibration Kits" dialog and
its various sub-dialogs.
Calibration kits and connector types are global resources; the parameters are stored
independently and are available irrespective of the current setup.
3.4.2 Calibration Types
The analyzer provides a wide range of calibration types for one, two or more ports. The
calibration types differ in the number and types of standards used, the error terms, i.e.
the type of systematic errors corrected and the general accuracy. The following table
gives an overview.
Table 3-2: Overview of calibration types
Calibration Type
Standards
Parameters
Error Terms
General Accuracy
Application
Reflection Normali-
zation
Open or Short
S
11
(or S
22
, ...)
Reflection tracking
Low to medium
Reflection measure-
ments on any port.
Transmission Nor-
malization
Through
S
12
, S
21
(or S
13
,...)
Transmission track-
ing
Medium
Transmission meas-
urements in any
direction and
between any combi-
nation of ports.
Full One-Port
Open, Short,
Match
1)
S
11
(or S
22
, ...)
Reflection tracking,
Source match
Directivity,
High
Reflection measure-
ments on any port.
One-Path Two-Port
Open, Short,
Match
1)
(at source
port),
Through
2)
S
11
, S
21
(or S
22
,...)
Reflection tracking,
Source match,
Directivity,
Transmission track-
ing
Medium to high
Unidirectional trans-
mission measure-
ments in any direc-
tion and between
any combination of
ports.
TOSM (2-port, 3-
port or 4-port) or
UOSM
Open, Short,
Match
1)
(at each
port),
Through
2)
(between
all combinations of 2
ports)
All
Reflection tracking,
Source match,
Directivity,
Load match,
Transmission track-
ing,
High
Reflection and
transmission meas-
urements on DUTs
with 2, 3, or 4 ports;
classical 12-term
error correction
model.
TOM (2-port, 3-port
or 4-port)
Open, Match (at
both ports),
Through
All
Reflection tracking,
Source match,
Directivity,
Load match,
Transmission track-
ing
High, implicit verifi-
cation
Reflection and
transmission meas-
urements on DUTs
with 2, 3, or 4 ports.
Calibration