Remote Commands
R&S
®
Spectrum Rider FPH
365
User Manual 1321.1011.02 ─ 06.00
[SENSe:]PMETer:FREQuency
<Frequency>
This command sets the frequency of the power sensor.
Parameters:
<Frequency>
Specified in the data sheet.
Default unit: GHz
Example:
PMET:FREQ 500 MHZ
Sets the power sensor's frequency to 500 MHz.
CALCulate<1...2>:PMETer:CPOWer:BANDwidth
<ChannelBW>
This command defines the channel bandwidth.
Available for the channel power meter.
The numeric suffix at CALCulate is irrelevant for this command.
Parameters:
<ChannelBW>
Default unit: GHz
Example:
CALC:PMET:CPOW:BAND 5 MHZ
Sets the channel bandwidth to 5 MHz.
13.6.4.2
Configuring Power Level Readout
The following chapter describes commands that configure the power level readout.
List of commands
●
CALCulate<1...2>:PMETer:RELative[:MAGNitude]
●
CALCulate<1...2>:PMETer:RELative[:MAGNitude]:AUTO
●
CALCulate<1...2>:PMETer:RELative[:MAGNitude]:OFFSet
●
CALCulate<1...2>:PMETer:RELative[:MAGNitude]
<RefLevel>
This command sets the reference value for relative measurements.
The numeric suffix at CALCulate is irrelevant for this command.
Parameters:
<RefLevel>
Default unit: dBm
Example:
CALC:PMET:REL 30
The reference value to 30 dBm.
CALCulate<1...2>:PMETer:RELative[:MAGNitude]:AUTO
<ONCE>
This command sets the current measurement result as the reference level for relative
measurements.
Remote Control - Commands