Remote Commands
R&S
®
Spectrum Rider FPH
410
User Manual 1321.1011.02 ─ 06.00
larly since faulty commands from the controller to the instrument are recorded there as
well.
13.6.11.6
Reset Values of the Status Reporting System
contains the different commands and events causing the status reporting
system to be reset. None of the commands, except
*RST
and
SYSTem:PRESet
, influ-
ences the functional instrument settings. In particular, DCL does not change the instru-
ment settings.
Table 13-13: Resetting the status reporting system
Event
Switching
on supply
voltage
DCL, SDC
Power-On-
Status-Clear
(Device
Clear,
Selected
Device
Clear)
*RST or
SYS-
Tem:PRE-
Set
STA-
Tus:PRESet
*CLS
Effect
0
1
Clear STB, ESR
-
yes
-
-
-
yes
Clear SRE, ESE
-
yes
-
-
-
-
Clear PPE
-
yes
-
-
-
-
Clear EVENt parts of
the registers
-
yes
-
-
-
yes
Clear ENABle parts
of all OPERation and
QUEStionable regis-
ters; Fill ENABle
parts of all other reg-
isters with "1".
-
yes
-
-
yes
-
Fill PTRansition parts
with "1"; Clear
NTRansition parts
-
yes
-
-
yes
-
Clear error queue
yes
yes
-
-
-
yes
Clear output buffer
yes
yes
yes
1)
1)
1)
Clear command pro-
cessing and input
buffer
yes
yes
yes
-
-
-
1) Every command being the first in a program message, i.e., immediately following a
<PROGRAM MESSAGE TERMINATOR> clears the output buffer.
13.6.11.7
Remote Commands of the Status Reporting System
The following commands control the status-reporting system
*RST
does not influence
the status registers.
Remote Control - Commands