Spectrum Analyzer Mode
R&S
®
Spectrum Rider FPH
152
User Manual 1321.1011.02 ─ 06.00
Figure 6-9: Long time recording settings dialog box
Long time recording
settings
Descriptions
"Recording Mode"
There are three selections for the recording mode: "Timer", "Immediately"
and "Limits Failure"
When "Long Time Recording Settings" is enabled with recording mode
"Timer", the configured "Start Date"/ "Start Time" and "Stop Date"/ "Stop
Time" and "Duration" determine the time window in which the spectrogram
data is recorded. The setting for "Limits Save Mode" is disabled.
When "Long Time Recording Settings" is enabled with recording mode "Start
Immediately", spectrogram data recording starts immediately and "Duration"
determines when the recording of the spectrogram data stopped. The set-
tings of "Start Date", "Start Time" and "Stop Date" and "Stop Time" is dis-
abled.
When "Long Time Recording Settings" is enabled with recording mode "Lim-
its Failure", the "Limits Save mode" determines when data is recorded. The
settings for "Start Date", "Start Time" and "Stop Date" and "Stop Time" is dis-
abled.
By default, "Timer" is selected.
"Recording Speed"
There are two selections for the recording speed: "Max" and "Manual".
If recording speed is set to "Manual", the recording interval is configurable.
By default, "Max" is selected.
"Start Date"
Start date format: dd/mm/yy.
"Stop Date"
Stop date format: dd/mm/yy.
"Storage"
Two storage types are supported: "Storage Card" and "USB".
By default, "USB" is selected.
Performing Spectrum Measurements