Remote Commands
R&S
®
Spectrum Rider FPH
366
User Manual 1321.1011.02 ─ 06.00
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command.
Parameters:
<ONCE>
ONCE
Example:
CALC:PMET:REL ONCE
CALCulate<1...2>:PMETer:RELative[:MAGNitude]:OFFSet
<RefLvlOffset>
This command sets an offset for the reference value.
The numeric suffix at CALCulate is irrelevant for this command.
Parameters:
<RefLvlOffset>
Default unit: dB
Example:
CALC:PMET:REL -10
UNIT<1...2>:PMETer:POWer
<Unit>
This command selects the unit of the power sensor.
The numeric suffix at UNIT has the following effects:
Table 13-4: Power measurement with R&S FSH-Z1, R&S FSH-Z18 and USB power sensors:
Unit 1
Power unit.
Unit 2
Not available.
Table 13-5: Power measurement with R&S FSH-Z14 and R&S FSH-Z44:
Unit 1
Forward power.
Unit 2
Reflected power.
Suffix:
<1...2>
.
1...2
Parameters:
<Unit>
DBM | DB | WATT | VSWR | W | DUVM | DBUV | W_M2 | V_M
Note on the parameter DB: when applied to UNIT1, the power is
relative to the reference level, when applied to UNIT2, the return
loss is displayed.
Note on the parameter VSWR: the parameter is only available if
applied to UNIT2.
Example:
UNIT1:PMET:POW DBM
When measuring with the R&S FSH-Z1, R&S FSH-Z18 or USB
power sensors: sets unit to dBm.
When measuring with the R&S FSH-Z14 or R&S FSH-Z44: sets
unit of forward power to dBm.
Remote Control - Commands