Suffix:
<t>
.
<range>
1...3
Selects the user range.
Example:
FETC:PNO:USER2:RMS?
Queries the RMS jitter for user range 2.
Usage:
Query only
Manual operation:
See
FETCh:PNOise<t>:USER<range>:RPM?
This command queries the residual PM for a particular user range.
The trace that is queried depends on
CALCulate<n>:EVALuation:USER<range>:
.
Suffix:
<t>
.
<range>
1...3
Selects the user range.
Example:
FETC:PNO:USER2:RPM?
Queries the residual PM for user range 2.
Usage:
Query only
Manual operation:
See
8.11.3 Configuring Spot Noise Measurements
......................................................................................... 131
CALCulate<n>:SNOise:DECades[:STATe]
....................................................................... 132
CALCulate<n>:SNOise:DECades:X?
............................................................................... 132
CALCulate<n>:SNOise:DECades:Y?
............................................................................... 132
................................................................................. 133
......................................................................................... 133
........................................................................................134
DISPlay[:WINDow<n>]:TRACe<t>:SELect
........................................................................134
CALCulate<n>:SNOise:AOFF
This command turns all spot noise markers off (custom and 10
x
markers).
Suffix:
<n>
.
Example:
CALC:SNO:AOFF
Turns all spot noise markers off.
Usage:
Event