Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
150
User Manual 1173.9286.02 ─ 15
8.15.1.3
STATus:QUEStionable:LIMit Register
The
STATus:QUEStionable:LIMit
register contains information about limit lines
and the results of a limit checks.
The number of LIMit registers depends on the number of measurement windows avail-
able in any application.
Bit no
Meaning
0
LIMit 1 FAIL
This bit is set if limit line 1 is violated.
1
LIMit 2 FAIL
This bit is set if limit line 2 is violated.
2
LIMit 3 FAIL
This bit is set if limit line 3 is violated.
3
LIMit 4 FAIL
This bit is set if limit line 4 is violated.
4
LIMit 5 FAIL
This bit is set if limit line 5 is violated.
5
LIMit 6 FAIL
This bit is set if limit line 6 is violated.
6
LIMit 7 FAIL
This bit is set if limit line 7 is violated.
7
LIMit 8 FAIL
This bit is set if limit line 8 is violated.
8
Phase Noise LIMit FAIL
This bit is set if a limit of the phase noise limit line is violated.
9 to 14
Unavailable for phase noise measurements.
15
This bit is always 0.
8.15.1.4
STATus:QUEStionable:PNOise Register
The
STATus:QUEStionable:PNOise
register contains information about the status
of phase noise measurements.
Bit no
Meaning
0
No trace are active
This bit is set if no trace is on.
1
SIGNal not found
This bit is set if no valid signal could be found.
2
VERify
This bit is set if signal verification has failed.
Using the Status Register