Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
145
User Manual 1173.9286.02 ─ 15
8.12.4 Using the Marker Zoom
CALCulate:MARKer:FUNCtion:ZOOM
..............................................................................145
..................................................................................145
CALCulate:MARKer:FUNCtion:ZOOM
<ZoomFactor>
This command defines the factor or magnitude of the marker zoom.
Parameters:
<ZoomFactor>
Range:
1 to 20
*RST:
Zoom OFF
Example:
CALC:MARK:FUNC:ZOOM 12
Zooms into the diagram with a factor of 10 around marker 1.
Manual operation:
See
DISPlay[:WINDow:]ZOOM[:STATe]
<State>
This command turns the marker zoom for marker 1 on and off.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
DISP:ZOOM ON
Turns on the marker zoom.
Manual operation:
See
8.13 Configuring In- and Outputs
The following commands are necessary to configure the frontend settings.
INPut:CONNector
<ConnType>
Determines whether the RF input data is taken from the RF input connector or the
optional Analog Baseband I connector. This command is only available if the Analog
Baseband interface (R&S
FSW-B71) is installed and active for input. It is not available
for the R&S
FSW67 or R&S
FSW85.
Parameters:
<ConnType>
RF
RF input connector
AIQI
Analog Baseband I connector
*RST:
RF
Configuring In- and Outputs