Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
101
User Manual 1173.9286.02 ─ 15
Parameters:
<Mode>
IQ
Measurement based on I/Q data.
NORMal
Measurement based on spectrum analyzer data.
FFT
Measurement based on spectrum analyzer data. Kept for com-
patibility to R&S
FSV.
*RST:
Depends on half decade
Example:
BAND:TYPE FFT
Selects FFT analysis for all half decades.
[SENSe:]FREQuency:STARt
<Frequency>
This command defines the start frequency of the measurement range.
Parameters:
<Frequency>
Offset frequencies in half decade steps.
Range:
1 Hz to 3 GHz
*RST:
1 kHz
Example:
FREQ:STAR 10kHz
Defines a start frequency of 10 kHz.
Manual operation:
See
[SENSe:]FREQuency:STOP
<Frequency>
This command defines the stop frequency of the measurement range.
Parameters:
<Frequency>
Offset frequencies in half decade steps.
Range:
3 Hz to 10 GHz
*RST:
1 MHz
Example:
FREQ:STOP 10MHz
Defines a stop frequency of 10 MHz.
Manual operation:
See
[SENSe:]LIST:BWIDth:RESolution:TYPE
<Mode>
This command selects the sweep mode for all half decades.
Configuring the Measurement Range