Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
81
User Manual 1173.9286.02 ─ 15
Example:
ABOR;*WAI
INIT:IMM
Aborts the current measurement and starts a new one once
abortion has been completed.
Usage:
Event
CONFigure:REFMeas ONCE
This command initiates a reference measurement that determines the inherent phase
noise of the R&S
FSW.
Parameters:
ONCE
Example:
CONF:REFM ONCE
Initiates a reference measurement
Manual operation:
See
INITiate<n>:CONMeas
This command restarts a (single) measurement that has been stopped (using
ABORt
)
or finished in single sweep mode.
The measurement is restarted at the beginning, not where the previous measurement
was stopped.
, this command does not reset traces in
maxhold, minhold or average mode. Therefore it can be used to continue measure-
ments using maxhold or averaging functions.
Suffix:
<n>
.
irrelevant
Usage:
Event
Manual operation:
See
INITiate<n>:CONTinuous
<State>
This command controls the sweep mode for an individual measurement channel.
Note that in single sweep mode, you can synchronize to the end of the measurement
with *OPC, *OPC? or *WAI. In continuous sweep mode, synchronization to the end of
the measurement is not possible. Thus, it is not recommended that you use continuous
sweep mode in remote control, as results like trace data or markers are only valid after
a single sweep end synchronization.
For details on synchronization see the "Remote Basics" chapter in the R&S
FSW User
Manual.
Suffix:
<n>
.
irrelevant
Performing Measurements