Analysis
R&S
®
FSW-K40
53
User Manual 1173.9286.02 ─ 15
Remote command:
Top:
DISPlay[:WINDow]:TRACe:Y[:SCALe]:RLEVel
Range:
DISPlay[:WINDow]:TRACe:Y:[:SCALe]
Bottom:
DISPlay[:WINDow]:TRACe:Y[:SCALe]:RLEVel:LOWer
Auto Scale Once
Automatically scales the vertical axis for ideal viewing.
6.1.2 Configuring Traces
....................................................................................56
Frequency Drift: Trace 1 (On Off)
................................................................................. 56
Spectrum Monitor: Raw Trace / Trk Trace (On Off)
...................................................... 56
Trace Offset
Defines a trace offset in dB.
The trace offset moves the trace vertically by the level you have defined.
The range is from -200
dB to 200
dB.
Remote command:
DISPlay[:WINDow]:TRACe:Y[:SCALe]:RLEVel:OFFSet
Trace Smoothing
Defines the magnitude (or aperture) of trace smoothing in percent.
The range is from 1% to 20%. The aperture takes effect on all traces that you smooth.
For more information see
Chapter 4.5.3, "Trace Smoothing"
Remote command:
DISPlay[:WINDow]:TRACe<t>:SMOothing:APERture
Smoothing Type
Selects the method that the application uses to smooth the trace.
For more information see
Chapter 4.5.3, "Trace Smoothing"
"Linear"
Converts the data to linear values before smoothing the trace.
"Logarithmic"
Smoothes the (original) logarithmic data.
Configuring Graphical Result Displays