Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
123
User Manual 1173.9286.02 ─ 15
Manual operation:
See
DISPlay[:WINDow]:TRACe:Y[:SCALe]:RLEVel:OFFSet
<Offset>
This command defines the trace offset.
Parameters:
<Offset>
Range:
-200 to 200
*RST:
0dB
Default unit: dB
Example:
DISP:TRAC:Y:RLEV:OFFS -10
Defines a trace offset of -10 dB.
Manual operation:
See
FORMat:DEXPort:DSEParator
<Separator>
This command selects the decimal separator for data exported in ASCII format.
Parameters:
<Separator>
COMMa
Uses a comma as decimal separator, e.g.
4,05
.
POINt
Uses a point as decimal separator, e.g.
4.05
.
*RST:
*RST has no effect on the decimal separator.
Default is POINt.
Example:
FORM:DEXP:DSEP POIN
Sets the decimal point as separator.
Manual operation:
See
FORMat:DEXPort:HEADer
<State>
If enabled, additional instrument and measurement settings are included in the header
of the export file for result data. If disabled, only the pure result data from the selected
traces and tables is exported.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Manual operation:
See
FORMat:DEXPort:TRACes
<Selection>
This command selects the data to be included in a data export file (see
Graphical Display of Phase Noise Results