![Rohde & Schwarz FSW-K40 User Manual Download Page 141](http://html.mh-extra.com/html/rohde-and-schwarz/fsw-k40/fsw-k40_user-manual_1477846141.webp)
Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
141
User Manual 1173.9286.02 ─ 15
CALCulate<n>:MARKer<m>:Y?
This command queries the position of a marker on the y-axis.
If necessary, the command activates the marker first.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Return values:
<Level>
Phase noise level of the marker.
Default unit: dBc/Hz
Example:
INIT:CONT OFF
Switches to single measurement mode.
CALC:MARK2 ON
Switches marker 2.
INIT;*WAI
Starts a measurement and waits for the end.
CALC:MARK2:Y?
Outputs the measured value of marker 2.
Usage:
Query only
8.12.2 Using Delta Markers
CALCulate<n>:DELTamarker<m>:AOFF
..........................................................................141
CALCulate<n>:DELTamarker<m>:MREFerence
............................................................... 142
CALCulate<n>:DELTamarker<m>[:STATe]
.......................................................................142
CALCulate<n>:DELTamarker<m>:TRACe
........................................................................142
................................................................................ 143
CALCulate<n>:DELTamarker<m>:Y?
...............................................................................143
CALCulate<n>:DELTamarker<m>:AOFF
This command turns off
all
delta markers.
Suffix:
<n>
.
<m>
irrelevant
Example:
CALC:DELT:AOFF
Turns off all delta markers.
Usage:
Event
Using Markers