Configuration
R&S
®
FSW-K40
44
User Manual 1173.9286.02 ─ 15
Range Start / Stop
Defines the frequency offsets that make up the measurement range.
Note that the maximum offset you can select depends on the hardware you are using.
on page 39 is used, make sure the entire frequency range is cov-
ered by the preamplifier.
Remote command:
Measurement Range Start
Measurement Range Stop
Sweep Forward
Selects the sweep direction. Forward and reverse sweep direction are available.
Forward sweep direction performs a measurement that begins at the smallest fre-
quency offset you have defined. The measurement ends after the largest offset has
been reached.
Reverse sweep direction performs a measurement that begins at the largest frequency
offset you have defined. The measurement ends after the smallest offset has been
reached. The reverse sweep is the default sweep direction because the application is
able to lock on a drifting carrier frequency in that case.
Remote command:
Presets
Selects predefined measurement settings for each individual half decade that are used
for the measurement.
If the measurement settings differ from one of the preset states, the application dis-
plays a symbol ( ) at the label.
"Fast"
Fast measurements perform one measurement in each half decade.
No averaging takes place.
"Normal"
Normal measurements use averaging for some half decades, but with
respect to measurement speed.
"Average"
Average measurements use averaging for all half decades. However,
you have to put up with slower measurement speed.
"Manual"
Manual configuration of the measurement range.
Remote command:
Global RBW
Defines the resolution bandwidth for all half decades globally.
The resulting RBW is a percentage of the start frequency of the corresponding half
decade.
If the resulting RBW is not available, the application rounds to the next available band-
width.
Configuring the Measurement Range