Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
122
User Manual 1173.9286.02 ─ 15
DISPlay[:WINDow]:TRACe:Y[:SCALe]:MANual
<Mode>
This command selects the type of manual scaling of the vertical axis.
Parameters:
<Mode>
BRANge
Scaling based on the value at the bottom of the diagram and the
axis range.
OFF
Turns manual scaling of the y-axis off.
TBOTtom
Scaling based on the values on the bottom and top of the dia-
gram.
TRANge
Scaling based on the value at the top of the diagram and the
axis range.
*RST:
ON
Example:
DISP:WIND2:TRAC:Y:SCAL:MAN TRAN
Scaling of the vertical axis based on the top and the range of the
axis.
Manual operation:
See
DISPlay[:WINDow]:TRACe:Y[:SCALe]:RLEVel
<Reference>
This command defines the reference value or upper border of the diagram area.
Note that you have to select manual y-axis scaling before you can use the command.
Parameters:
<Reference>
Range:
-200 to 0
*RST:
-20
Default unit: dBc/Hz
Example:
DISP:TRAC:Y:RLEV -50
Defines a reference value of -50 dBc/Hz.
Manual operation:
See
DISPlay[:WINDow]:TRACe:Y[:SCALe]:RLEVel:LOWer
<LowerReference>
This command defines the reference value or upper border of the diagram area.
Note that you have to select manual y-axis scaling before you can use the command.
Parameters:
<LowerReference>
Range:
-400 to 1
*RST:
-120
Default unit: dBc/Hz
Example:
DISP:TRAC:Y:RLEV:LOW -100
Sets the bottom of the diagram to .100 dBc/Hz.
Graphical Display of Phase Noise Results