Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
104
User Manual 1173.9286.02 ─ 15
Manual operation:
See
"Half Decades Configuration Table"
[SENSe:]LIST:RANGe<range>:SWEep:COUNt
<Measurements>
This command defines the number of measurements included in the averaging for a
half decade.
Suffix:
<range>
.
1...20
Selects the half decade.
For the suffix assignment see
.
Parameters:
<Measurements>
Range:
1 to 10000
*RST:
Depends on the half decade
Example:
LIST:RANG9:SWE:COUN 15
Includes 15 measurements in the averaging of the ninth half
decade.
Manual operation:
See
"Half Decades Configuration Table"
[SENSe:]LIST:SWEep:COUNt
<Averages>
This command defines the number of measurements to be included in the averaging
for each and all half decades.
Parameters:
<Averages>
Range:
1 to 10000
*RST:
Depends on the half decade.
Example:
LIST:SWE:COUN 20
Defines 20 averages for all half decades.
Manual operation:
See
[SENSe:]LIST:SWEep:COUNt:MULTiplier
<Multiplier>
This command defines a multiplier that is applied to the average count in each half
decade.
Before you can use the command you have to turn on the multiplier with
]LIST:SWEep:COUNt:MULTiplier:STATe
.
Parameters:
<Multiplier>
Numeric value.
Example:
LIST:SWE:COUN:MULT:STAT ON
LIST:SWE:COUN:MULT 5
Turns on the multiplier and multiplies the averages by 5.
Manual operation:
See
Configuring the Measurement Range