Configuration
R&S
®
FSW-K40
48
User Manual 1173.9286.02 ─ 15
If the trace modes "Average" , "Max Hold" or "Min Hold" are set, this value also deter-
mines the number of averaging or maximum search procedures.
In continuous sweep mode, if "Sweep Count" = 0 (default), averaging is performed
over 10 sweeps. For "Sweep Count" =1, no averaging, maxhold or minhold operations
are performed.
Remote command:
Finish Half Decade
Aborts the measurement in the current half decade and continues measuring the sub-
sequent half decade.
Averaged results displayed for a half decade finished prematurely are based on the
number of measurements already done.
Remote command:
5.7 Configuring In- and Outputs
The "In- / Output" menu contains all functionality necessary to control and perform
phase noise measurements.
The R&S FSW Phase Noise Measurement application can also process input from an
external mixer. For details on configuring the external mixer see the manual of the
R&S
FSW.
..................................................................................... 48
5.7.1 Input Source Configuration
Access
: "Overview" > "Input / Frontend" > "Input Source Config"
The "Radio Frequency" dialog box contains functionality to configure the input source.
Note that the "Radio Frequency (On Off)" button is unavailable in the R&S
FSW-K40
Coupling
Selects the coupling method at the RF input.
AC coupling blocks any DC voltage from the input signal. DC coupling lets DC voltage
through.
For more information refer to the data sheet.
Remote command:
Configuring In- and Outputs