How to Configure Phase Noise Measurements
R&S
®
FSW-K40
68
User Manual 1173.9286.02 ─ 15
7 How to Configure Phase Noise Measure-
ments
7.1 Performing a Basic Phase Noise Measurement
1. In the Spectrum application, define the center frequency of the DUT.
2. Enter the "Phase Noise" application.
The R&S
FSW-K40 starts the measurement with the default configuration. The
default configuration defines most settings automatically.
If you need any custom configuration, define them after entering the Phase Noise
application.
3. Layout the display as required via the SmartGrid.
4. Open the "Overview" dialog box to configure the measurement.
5. Configure the frontend (frequency, level etc.) via the "Frontend" dialog box.
6. Define the measurement range via the "Phase Noise" dialog box.
7. Turn on frequency and level tracking via the "Control" dialog box.
8. Run a single sweep.
9. Turn on a marker and read out the results.
10. Read out the residual noise over the measurement range.
11. Customize a residual noise range and read out the results.
12. Freeze trace 1 and 2 (trace mode: View).
13. Turn on trace 3 and 4 (trace mode: Clear/Write).
14. Switch the measurement mode to "IQ FFT" in the "Phase Noise" dialog box.
15. Repeat the measurement.
7.2 Customizing the Measurement Range
The application provides several ways to customize. Each method features a different
level of details you can define.
1. Open the "Phase Noise" configuration via the "Overview" dialog box or the "Meas
Config" softkey menu.
2. Define the frequency offset range you'd like to measure in the corresponding fields.
Customizing the Measurement Range