Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
99
User Manual 1173.9286.02 ─ 15
Manual operation:
See
[SENSe:]SWEep:SVFailed
<State>
This command turns repeated tries to start the measurement if signal verification fails
on and off.
Parameters:
<State>
ON | OFF | 1 | 0
If on, the application tries to verify the signal once and then
aborts the measurement if verification has failed.
*RST:
0
Example:
SWE:SVF ON
Stops the measurement if signal verification has failed.
Manual operation:
See
8.8 Configuring the Measurement Range
The following commands are necessary to configure the phase noise measurement
range.
Table 8-4: <range> suffix assignment
Suffix
Half Decade
1
1 Hz ... 3 Hz
2
3 Hz ... 10 Hz
3
10 Hz ... 30 Hz
4
30 Hz ... 100 Hz
5
100 Hz ... 300 Hz
6
300 Hz ... 1 kHz
7
1 kHz ... 3 kHz
8
3 kHz ... 10 kHz
9
10 kHz ... 30 kHz
10
30 kHz ... 100 kHz
11
100 kHz ... 300 kHz
12
300 kHz ... 1 MHz
13
1 MHz ... 3 MHz
14
3 MHz ... 10 MHz
15
10 MHz ... 30 MHz
16
30 MHz ... 100 MHz
Configuring the Measurement Range