Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
128
User Manual 1173.9286.02 ─ 15
Parameters:
<OffsetFrequency>
The minimum offset is 1
Hz. The maximum offset depends on
the hardware you are using.
*RST:
1 MHz
Example:
CALC:EVAL:USER2:STAR 1khz
Defines a start point of 1 kHz for the second user range.
Manual operation:
See
CALCulate<n>:EVALuation:USER<range>:STOP
<OffsetFrequency>
This command defines the end point of a custom residual noise calculation range.
Before you can use the command, you have to assign the user range to a trace with
CALCulate<n>:EVALuation:USER<range>:TRACe
Suffix:
<n>
.
Parameters:
<OffsetFrequency>
The minimum offset is 1
Hz. The maximum offset depends on
the hardware you are using.
*RST:
1 MHz
Example:
CALC:EVAL:USER2:STOP 100kHz
Defines an end point of 100 kHz for the second user range.
Manual operation:
See
CALCulate<n>:EVALuation:USER<range>:TRACe
<Trace>
This command selects the trace for a custom residual noise calculation range.
Suffix:
<n>
.
<range>
1...3
Selects the user range.
Parameters:
<Trace>
NONE
Turns a user range off.
TRACE1 ... TRACE6
Trace to assign the user range to.
Example:
CALC:EVAL:USER2:TRAC TRACE2
Assigns the second user range to trace 2.
Manual operation:
See
Configure Numerical Result Displays