R&S FSH
Spectrum Analyzer Mode
Performing Spectrum Measurements
Operating Manual 1309.6275.12 - 14
63
C
Freq [Hz]
C
Power Abs
Absolute peak power in the corresponding frequency range.
C
Power Rel
Relative peak power in relation to the channel power of the reference channel.
C
U
Limit
Minimum distance from the limit line to the trace in the corresponding frequency
range. Negative values or a zero indicate a passed SEM limit check, positive
values indicate a violation of the limit check.
2.1.6 Measuring the Harmonic Distortion
The Harmonic Distortion measurement is an easy way to identify the harmonics of a
DUT. In addition to the graphic display of the harmonics, this measurement mode also
calculates the Total Harmonic Distortion (THD) and shows the results.
You can perform a Harmonic Distortion measurement in frequency sweep (span > 0)
and zero span mode (span = 0). When starting the measurement, the R&S FSH looks
for the first harmonic of the signal (= the highest signal) in the defined frequency range.
It then adjusts the frequency axis so that all harmonics are visible. In zero span mode,
the center frequency remains the same.
M
Press the MEAS key.
M
Press the "Meas Mode" softkey.
M
Select the "Harmonic Distortion" menu item.
The R&S FSH starts to measure the harmonic distortion.
The search for harmonics starts as soon as you enter the Harmonic Distortion
measurement. Upon entering the measurement, the R&S FSH automatically adjusts
the settings in order to display the selected number of harmonics (default = 2) on the
screen.