R&S FSH
Receiver Mode
Configuring Measurements in Receiver Mode
Operating Manual 1309.6275.12 - 14
170
6.2.3 Defining the Measurement Time
The measurement time is the time that the R&S FSH collects data at each
measurement frequency to calculate the results for that frequency according to the
detector you have selected.
You can define a measurement time between 5 ms and 1000 s.
M
Press the SWEEP key.
M
Press the "Meas Time" softkey.
M
Define the measurement time you need.
Selecting the scan mode
When you enter the receiver mode, the R&S FSH repeatedly measures the receive
frequency or set of frequencies over the measurement time you have defined. If you
want to perform a single measurement or scan only, select single scan mode.
M
Press the SWEEP key.
M
Press the "Single Scan" softkey to perform a single scan or measurement.
M
Press the "Cont Scan" softkey to perform a continuous scan or measurement.
When you select single scan mode, the R&S FSH performs the measurement once
over the measurement time and then stops. In case of frequency scans, the
R&S FSH performs one measurement on each receive frequency in the scan
range over the measurement time and stops when it has measured all frequencies
that are part of the scan range.
6.2.4 Using Transducers