R&S FSH
Network Analyzer Mode
Performing Scalar Measurements
Operating Manual 1309.6275.12 - 14
129
Some measurement setups may include additional cables or adapters that may have
an additional electrical length. To correct their phase error, you can include the
electrical length of additional test equipment. The R&S FSH distinguishes if the
equipment is connected to port 1 or port 2.
M
Press the MEAS key.
M
Press the "Calibration" softkey.
M
Select the "Offset Length Port 1" or "Offset Length Port 2" menu item.
The R&S FSH opens an input field to define the electrical length of the accessory.
M
Enter the electrical length.
The R&S FSH now includes the electrical length of the accessory for phase
measurements.
4.2 Performing Scalar Measurements
With scalar measurements, you can measure the magnitude of both reflection and
transmission characteristics of a device under test. To calibrate the measurements,
only normalization is available.
4.2.1 Measuring the Transmission
The example included here is based on a transmission measurement of a two-port
filter. The filter has a center frequency of 2060 MHz and a bandwidth of about 11 MHz.
Test setup
M
Connect the input of the DUT to the tracking generator output (port 2).
M
Connect the output of the DUT to the RF input (port 1).
Preset the R&S FSH
Before starting the measurement procedure, preset the R&S FSH to restore the default
configuration and connect the filter between the measurement ports.
M
Press the PRESET key.
M
Connect the DUT.
Start scalar measurements
M
Press the MEAS key.
M
Press the "Meas Mode" softkey.
M
Select the "Scalar" menu item.