R&S FSH
Working with Power Sensors
Using a Power Sensor
Operating Manual 1309.6275.12 - 14
113
Screen layout of the power meter mode
1
Connected power sensor model
2
Reference for relative power measurements
3
Power offset
4
Measurement time
5
Readout of the measured power
6
Analog readout of the measured power
7
Measurement frequency
8
Power sensor softkey menu
3.1.1 Connecting a Power Sensor
The R&S FSH controls and powers the power sensors via a special interface on the
top of the instrument. You can also connect a power sensor via the USB interface on
the right.
If you are using the R&S FSH-Z1 and -Z18 power sensors, connect the power sensor
cable to the power sensor interface and screw it into position. Power sensors of the
R&S NRP product line are controlled via the USB interface with a passive USB
adapter.
After connecting the power sensor to the R&S FSH, you can connect the DUT to the N-
connector of the power sensor.
Risk of damage to the power sensor
Before you start to work with the power sensor, make sure that the continuous power
applied to the input of the power sensor does not exceed a certain level.
Refer to the documentation of the power sensor for more information on the maximum
input power.