R&S FSH
Distance-to-Fault Mode (R&S FSH-K41)
Configuring Cable and Antenna Tests
Operating Manual 1309.6275.12 - 14
156
5.2.1.1 Selecting a Predefined Cable Model
M
Press the MEAS key.
M
Press the "Cable Config" softkey.
M
Select the "Select Cable Model"
menu item.
The R&S FSH opens the file
manager to select the cable
model.
M
Select the cable model that you
are testing.
M
Confirm the selection with the
"Select" softkey.
The R&S FSH shows the currently selected cable model in the diagram header.
M
Select the "Deselect Cable Model" item from the "Cable Config" menu if you want
to perform measurements without using a particular cable model.
5.2.1.2 Creating a Cable Model
The R&S FSH provides two ways to define customized cable models.
The first way is to define a cable model with the "Cable Model Editor" that is part of the
R&S FSH4View software package. The R&S FSH4View is delivered with the
R&S FSH. With this software, you can define a cable model on a PC and then transfer
it to the R&S FSH. Then you can select it just like any other predefined cable model.
For more information see
C
"
Saving and Loading Instrument Settings and Measurement Results
If you do not have access to a PC, but still need a cable model that is not stored on the
R&S FSH, you can also define the characteristics of a cable temporarily on the
R&S FSH itself. It is not possible to save these in a data set, however. They are lost as
soon as you change them or load another cable model.
M
Press the MEAS key.
M
Press the "Cable Config" softkey.
M
Select the "Define User Model" menu item.
A submenu opens.
M
Select the "Frequency" menu item.
M
Enter the frequency of the cable you are testing.
M
Select the "Velocity" menu item.
M
Enter the cable velocity.
M
Select the "Loss" menu item.