R&S FSH
Spectrum Analyzer Mode
Performing Spectrum Measurements
Operating Manual 1309.6275.12 - 14
50
Screen layout for TDMA power measurements
1
Standard
2
TDMA power
3
Measurement limits
4
Trigger delay
5
Trigger level
6
Measurement time
7
TDMA power measurement softkey menu
2.1.3.1 Selecting a Standard
If you need to perform measurements that are conform to a telecommunications
standard, you can activate one of the predefined standards that are already stored in
the R&S FSH memory. When starting the measurement in the default configuration,
the R&S FSH automatically activates the GSM/EDGE standard. However, you can
create new configurations to perform measurements on other standards as well.
M
Press the MEAS key.
M
Press the "Standard" softkey.
The R&S FSH opens a dialog box to select the standard.
M
Select one of the available standards.
M
Confirm the selection with the "Select" softkey.
The R&S FSH loads the configuration of the selected standard. It automatically
sets the optimal settings for the selected standard.
You can create and edit standards with the R&S FSH4View software and transfer them
to the R&S FSH via the USB or the LAN interface. The number of standards the
R&S FSH can store in its memory depends on the number of other data sets stored on
the R&S FSH. For more information see "