R&S FSH
Distance-to-Fault Mode (R&S FSH-K41)
Vector Voltmeter (R&S FSH-K45)
Operating Manual 1309.6275.12 - 14
151
M
For measurement on DUTs that need an external voltage supply (e.g. power
amplifiers), you can connect the supply voltage from a suitable AC power supply to
the BIAS Port 1 or use the internal bias.
By default, the R&S FSH is in spectrum analyzer mode after you have turned it on. To
switch to the DTF mode from another operating mode, proceed as follows.
M
Press the MODE key.
M
Press the "Distance-to-Fault" softkey.
The R&S FSH starts the DTF mode.
By default, the R&S FSH performs measurements on port 1. It however, also supports
measurements on port 2, but you have to configure those first.
M
Press the "Meas Mode" softkey.
M
Select the "DTF / Refl Measurement Port" menu item.
The R&S FSH opens a submenu.
M
Select either "Port 1" or "Port 2".
Screen layout of the cable and antenna analyzer
1
Cable model
2
Hardware settings
3
Marker information
4
Status line:
- S-matrix
- Calibration status
- Measurement format
5
Trace window
6
Marker (blue line)
7
Cable length information
8
Cable frequency information
9
Cable test softkey menu