![Rohde & Schwarz FSH4 Operating Manual Download Page 195](http://html.mh-extra.com/html/rohde-and-schwarz/fsh4/fsh4_operating-manual_1477834195.webp)
R&S FSH
Digital Modulation Analyzer
Measurements on 3GPP FDD Signals
Operating Manual 1309.6275.12 - 14
176
7.2 Measurements on 3GPP FDD Signals
Equipped with the software application R&S FSH-K44, you can perform measurements
on WCDMA signals in accordance to the 3GPP standard with your R&S FSH.
You can expand the functionality by adding the R&S FSH-K44E firmware application.
This application performs code domain power measurements on downlink WCDMA
signals according to the standard.
M
Press the MODE key
M
Press the "Dig Mod Analyzer" softkey
M
Select "3GPP WCDMA BTS" menu item.
The R&S FSH starts the signal analysis.
When starting a measurement, the R&S FSH first records a section of the signal that
lasts 20 ms (or 2 WCDMA frames). From this information, it gets all necessary
information for further analysis of the WCDMA signal.
The R&S FSH performs various measurements on a WCDMA slot in general as well as
measurements on specific channels.
Measurements on slot level include the power level, error vector magnitude (EVM), the
code domain error and the frequency error. To get sufficient measurement accuracy, it
is necessary to synchronize the reference frequency of the base station and the
R&S FSH via the EXT REF IN input.
In addition, you can analyze the following channel types in more detail:
C
Common Pilot Channel (CPICH)
Note:
This channel is required by the channel configuration; without it,
synchronization is not possible.
C
Primary Common Control Physical Channel (P-CCPCH)
C
Primary Synchronization Channel (P-SCH)
C
Secondary Synchronization Channel (S-SCH)
For the CPICH and P-CCPCH channels the R&S FSH measures the power level and
the Ec/Io. For the P-SCH and S-SCH channels, the R&S FSH measures the power
level.
The 3GPP option provides several result displays to display the measurement results.
7.2.1 The Result Summary
The default result display is the result summary. The result summary shows various
measurement results and hardware settings in numerical form.
M
Press the "Result Display" softkey.
M
Select the "Result Summary" menu item.
The R&S FSH shows the numerical results in a table.