R&S FSH
Spectrum Analyzer Mode
Configuring Spectrum Measurements
Operating Manual 1309.6275.12 - 14
98
You can perform the following actions:
C
Position the marker with the cursor keys.
When positioning the marker with the cursor keys, the step size is 10% of the
span.
C
Position the marker with the rotary knob
When positioning the marker with the rotary knob, the step size is one pixel.
C
Enter a marker position with the number keys and confirm the entry with one of the
unit keys.
M
Confirm the marker position with the ENTER key.
The marker input field closes.
By default, the marker list above the diagram area is active. The marker list shows
the horizontal position of all markers and the corresponding vertical value. If
inactive, the list shows only the coordinates of markers 1 and 2.
You can turn it off and on again whenever you like.
M
Press the "View List" softkey.
The marker list turns off or on, depending on its original state.
2.2.6.2 Positioning a Delta Marker
When a normal marker is already in use, you can add delta markers.
M
Press the MARKER key.
M
Press the "New Marker" softkey.
The R&S FSH activates a delta marker and positions it on the next maximum level
that has been measured. In addition, the delta marker input field opens.
The R&S FSH adds the delta marker to the marker list and shows the marker
position relative to the normal marker (M1).
You can perform the following actions:
C
Enter a delta marker position with the number keys and confirm the entry with one
of the unit keys.
C
Change the delta marker position with the rotary knob or the cursor keys.
M
Confirm the delta marker position with the ENTER key.
The delta marker input field closes.
M
To add more markers, press the "New Marker" softkey several times until you have
the number of markers you want in the display.