Remote Control - OPC
R&S
®
QAR
103
User Manual 1178.7200.02 ─ 09
T
able 7-1: V
ariables for communication from PLC to R&S
QAR ("FromPLC")
Name
Type
Description
Common variables
Configuration
String
Measurement configuration (i.e. "G05-Base")
Type_Measurement
String
Measurement type.
●
"standard"
●
"reflection_normalization"
●
"transmission_normalization" (radome measurements only)
Config_Measurement
String
Application type.
●
"Default" (12 cluster application)
●
"SCluster" (1 cluster application)
Path_Results
String
Path to store measurement results (i.e. "C:\QAR\Results")
Path_Diagnostics
String
Path to store measurement results (i.e. "C:\QAR\Results")
Trigger.Measure
Bool
T
rigger to start measurement
Trigger.Diagnostics
Bool
T
rigger to start diagnostics
Heartbeat
Bool
Signal pulse to ensure communications
Results_Received
Bool
Signal to acknowledge reception of results
V
ariables for radome measurements
RadomeMeasurement.Reflection.Window.Width
Double
Reflection statistics window width.
RadomeMeasurement.Reflection.Window.Height
Double
Reflection statistics window height.
RadomeMeasurement.Reflection.Window.XCenter
Double
Reflection statistics window x center
.
RadomeMeasurement.Reflection.Window.YCenter
Double
Reflection statistics window y center
.
RadomeMeasurement.Transmission.Verification_Mode
Bool
V
erification mode.
OPC Variables