Remote Control - SCPI
R&S
®
QAR
92
User Manual 1178.7200.02 ─ 09
Return values:
<StopFrequency>
Default unit: GHz
Example:
//Query stop frequency of second frequency band
MEAS:TRAN:BAND2:STOP:FREQ?
Usage:
Query only
MEASurement:TRANsmission:MINFrequency?
This command queries the frequency at which the minimum transmission loss has
been measured.
Return values:
<Frequency>
Default unit: Hz
Example:
//Query frequency of the minimum transmission loss
MEAS:TRAN:MINF?
Usage:
Query only
MEASurement:TRANsmission:MINLoss?
This command queries the minimum transmission loss that has been measured.
Return values:
<TransmissionLoss> Default unit: dB
Example:
//Query the minimum transmission loss
MEAS:TRAN:MINL?
Usage:
Query only
MEASurement:TRANsmission:NORMalize:REQuired?
This command queries the normalization state of the transmission measurement.
Return values:
<State>
0
Measurement is normalized.
1
Measurement is not normalized. In that case, you should nor-
malize the measurement before starting it.
Example:
//Query normalization state
MEAS:TRAN:NORM:REQ?
Usage:
Query only
MEASurement:TRANsmission:NORMalize:STARt
This command initiates the normalization of a transmission measurement.
Remote Commands