Preparing for Use
R&S
®
QAR
28
User Manual 1178.7200.02 ─ 09
4. Move the reflector to the front or back as necessary.
5. Repeat normalization measurement.
a) If normalization does no longer indicate a misalignment, reassemble the stand.
b) If normalization still indicates a misalignment, repeat the adjustment until the
error no longer occurs.
Correcting a horizontal misalignment (z-axis)
Solves message in user interface: "Shift reference reflector downward / upward".
1. Loosen the screws on the side of the support bar with an Allen wrench, size 6 (2x).
2. Move the reflector up or down as necessary.
3. Repeat normalization measurement.
a) If normalization does no longer indicate a misalignment, reassemble the stand.
b) If normalization still indicates a misalignment, repeat the adjustment until the
error no longer occurs.
Correcting a horizontal tilt
Solves message in user interface: "Rod alignment or rod distance is not correct".
Installing the R&S
QAR