Reference Manual Rev 1.13
004R-646-113
Page 65
12.3.3.
CAL (Scale calibration)
Path
Description
SCALE
└ CAL
└ZERO
(
)
└SPAN
(
)
└ED.LIN
(
)
└CLR.LIN
(
)
└DIR.ZERO
(
)
└DIR.SPN
(
)
└DEF.CAL
(
)
Calibrate Scale
ZERO
: Perform a zero calibration.
SPAN
: Perform a span calibration. A zero
calibration should be done before doing a span
calibration.
ED.LIN
: Add or Modify linearization points.
CLR.LIN
: Clear unwanted linearization points.
DIR.ZER
(Direct mV/V Zero Calibration): Enter
signal strength (in mV/V) of zero calibration directly.
DIR.SPN
(direct mV/V span Calibration): Enter the
signal strength (in mV/V) of fullscale directly. No
test weights required.
DEF.CAL
(Default Calibration): Restore instrument
to default factory calibration and reset all items in
the SCALE menu to defaults.
: This item is trade critical and will affect the calibration counter(s) if changed.
12.3.4.
QA (QA alarm)
Path
Description
SCALE:
└ QA
└QA.OPT
(
)
└QA.YEAR
(
)
└QA.MONTH
(
)
└QA.DAY
(
)
Configure the quality assurance feature.
If active the instrument displays a ‘QA DUE’ warning
after the date limit has expired.
QA.OPT
: Turn QA feature on or off.
QA.YEAR, QA.MONTH, QA.DAY
: Enter QA expiry
date.
: This item is trade critical and will affect the
calibration counter(s) if changed.
QA.OPT Values
(
)
<OPT>
Off
(Default)
, On
QA.DATE Values
(
)
<NUM>
2000-01-01 To 2099-12-31